BECKMAN COLUTER. Streamlined, simplified separations.
Skip Steps in Cell Culturing Directly centrifuge your cell culture flasks
| |
| |
|
| UNSPSC: | 41103904 |
| Height: | 34.3 cm (13.5 in) |
| Width: | 76.2 cm (30 in) |
| Depth: | 62.2 cm (24.5 in) |
| Weight: | 121 kg (267 lb) |
| Approximate Acceleration Time: | 10 Acceleration rates |
| Approximate Deceleration Time: | 11 Deceleration rates |
| Drive Type: | Brushless Induction |
| Electrical Requirements: | 208 V, 60 Hz instrument, 187-229 VAC, 9 A |
| Max Speed: | 10,200 rpm with fixed-angle rotor; 3,750 rpm with swinging bucket rotor |
| Noise Level 0.91 meters (3 feet) in Front of Centrifuge: | = 68 dBa |
| Set Speed: | 200 to 10,200 rpm in 10-rpm increments |
| Set Temperature: | -10 to +40°C in 1° increments |
| Set Time: | Timed (Up to 99 hours, 59 min); Hold, Pulse (short run) |
| Temperature Control: | ± 2°C of set temperature (after equilibration) |
| Ambient Temperature Range: | 10 to 35°C |
| Humidity Restrictions: | < 75% (noncondensing) |
| Pollution Degree: | 2 |
| Audible End-of-Run Tone: | Yes |
| Automatic Rotor Identification: | Yes |
| Control panel material: | Coated polystyrene copolymer |
| Housing Surfaces: | Acrylic baking enamel |
| Installation Category: | Category II |
| Refrigeration: | Yes, CFC-Free |
| Relative Centrifugal Field (RCF): | 11,400 x g with fixed-angle rotor; 3,270 x g with swinging bucket rotor |
| Speed Display: | Actual rotor speed in 10-rpm increments |
| Timed Run: | Indicates run time remaining |
| User-Defined Programs: | 10 |
| Ventilation Clearances Required: | 7.6 cm (3.0 in) (sides and rear) |